This is a step-by-step description on how to set adjust the equalizer based on measurements in many positions. This can be applied to the sound systems/PA in for example movie theaters and underground stations.
A short description: Measurements are perfomed in a number of positions and averaging is applied to obtain a spatially averaged frequency response. We want to flatten this spatially averaged curve since a flat curve should sound better.
To make this as fast as possible, we don’t want to measure all 10 positions again if we change the EQ settings, but only perform one measurement.
To make this as simple as possible, we want WinMLS to show us a flat curve when the EQ is “correctly” tuned. (From this curve we also want to be able to read out the dB-value we need to adjust each band.)
To obtain this, this spatially average response is inverted and used as a “compensation” in the measurement system.
We also must do one of the following
• Either: Form a loop back with the equalizer only and perform measurements while adjusting the EQ settings until we smooth the inverted average response. The equalizer is then re-inserted into the system.
• Or: If you do not have access to the EQ output signal you can do this. Set the last measurement as reference. Perform measurements while adjusting the EQ settings until we smooth the inverted average response.
As result we have a reasonable smoothed average spatial frequency response.
Level calibration is not reqired for this application since it is based on comparing measurements. But if you are using a microphone that is not flat, this should be compensated.
1. Go to Setup->Load… to load a setup that plots the frequency response with correct settings, we recommend Sound System_Adjust Before Performance.
2. Make sure Plot->Overlay is checked.
3. Make sure the EQ of your sound system is turned off.
4. If your microphone needs compensation, make sure this compensation is turned on. To do this go to Plot->Plot Type Settings->Frequeny Response/Spectrum… or click F5. In this dialog box, select the compensation file for your microphone and turn on the compensation as shown below.
In the same dialog, make sure that no shift or normalization is turned on as shown below.
5.
Perform a number of measurements in different positions (measure with the
equalizer turned off at N different locations). Since Plot->Overlay
is checked, the curves will be plotted on top of each other and when
finished there will be N curves displayed. In the figure below
N=10.
6. Click
the button to display the
mean value on top of all the plotted curves. The average room response (average magnitude
response) is then calculated.
7. Click
the button to display the
average curve only. (If
not only the mean value is plotted, go to Plot->Overlay… and make sure
that is
selected.)
8. Now we want to save this inverted curve to use it later. Go to File->Export Data… and find the folder where WinMLS was installed. Then find the subfolder named Tune EQ. Make sure the settings are set as shown in the dialog box below.
9. To set this curve as reference, go to Plot->Plot Type Settings->Frequeny Response/Spectrum… or click F5. In this dialog box, select the file you just saved (Data0) and turn on the compensation as shown below.
If you do step 10 below, also make sure the microphone compensation is turned off .
If you instead do step 11 below, the microphone compensation should not be changed.
10. You can
either do this step or the next step: Form a loop-back with the equalizer
only. In other words, connect the measurement system to the loudspeaker only and
leave out the microphone and the loudspeakers.
(When we perform another
measurement of the loop-back with the equalisation applied , we then display the curve
which is the
equalizer response multiplied with the average room response .)
11.
Do this step if you have not done the previous step: If you don’t have
access to the equalizer output, make sure the microphone is in the last
measurement position. We will use the measurements in this position to tune the
equalizer. To do this correctly, we need to divide by the measurement performed
in this position. Find the Active Plot Toolbar and make sure it is set as shown
below.
If
you alternatively want to set this from a dialog box, go to Plot->Plot
Type Settings->Frequeny Response/Spectrum… or use Ctrl + F and
make sure this measurement is selected as shown in the figure below (in this
example Meas. 10 was the last measurement position that is therefore set
as reference)
Make sure that the check box is not checked.
(When we perform another measurement with the equalization applied , we display the curve which is the equalizer response multiplied by the average room response .
Alternative: This operation could have been done after step 5 instead of here. If this has been done, you must not use any reference measurement in this step so in this case make sure is selected. This will speed up the calculation somewhat.)
12. In the
previous steps the reference is set. Now it is time to clean up after this. Make
sure the buttons are turned off as shown in the figure.
To remove the measurements needed
to make the reference, click the button and select the settings shown in the figure below (this
will remove all the curves in the plot and remove the measurements from
memory)
13. Perform measurements and tune the EQ so the curve is as flat as possible. When the curve is reasonably flat, this procedure is finished.
14. When finished, you might want to reload the setup. Or alternatively turn the microphone compensation off or select the compensation file for your microphone. Do this from Plot->Plot Type Settings->Frequeny Response/Spectrum….